Spontaneous Pacemaker Lead Retraction and Coiling
[...]. Spontaneous Pacemaker Lead Retraction and Coiling. CTICases. Published February 10, 2017. https://str-case-explorer.pages.dev/c/bw6uw
Imaging findings
Serial radiographs demonstrate spontaneous migration and coiling of a right atrial pacemaker lead. While initially well-positioned, subsequent radiographs over a two-month period show the lead has retracted out of the atrium and has wound itself into multiple coils within the pulse generator pocket, while the right ventricular lead remains fixed.
Key takeaways
Spontaneous retraction and coiling of a pacemaker lead within its pocket is a classic manifestation of Twiddler's syndrome or its variants, often driven by patient manipulation of the pulse generator or repetitive arm movements. In some cases, anatomical changes such as gravity-induced displacement of breast tissue in females standing upright can pull on leads, emphasizing the need to secure both the pulse generator and leads properly at the time of surgical implantation.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- Buffalo Chest (Bison Chest) Post-SternotomyArtifact
- Levoatrial Cardinal Vein in Turner SyndromeCongenital
- Granulomatosis with Polyangiitis (GPA) presenting as Diffuse Alveolar HemorrhageAutoimmune
- Granulomatosis with Polyangiitis (GPA) with Tracheobronchial and Nodular InvolvementAutoimmune
- Uncorrected D-Transposition of the Great ArteriesCongenital
- Congenitally Corrected Transposition of the Great Arteries (L-TGA) with Situs InversusCongenital
- Unilateral Pulmonary Vein Occlusion with Contralateral Pulmonary Arterial HypertensionPulmonary Vascular
- Dislodged Atrial Septal Defect (ASD) Occluder DeviceDevices & Post-op
See all cases from February 10, 2017 →
Related Devices & Post-op cases
- Watchman Left Atrial Appendage Closure Device
- Pacemaker Lead Dislodgement into the Inferior Vena Cava Post-Resuscitation
- Migrated IVC Filter
- Protek Duo Device for Right Ventricular Support Post-LVAD
- Iatrogenic RV Perforation with Hemopericardium from Retained Thrombolysis Catheter
- HeartMate 3 Left Ventricular Assist Device (LVAD)