Myocardial Perforation by Pacemaker Lead with Insulation Disruption
▶ Watch on YouTube — age-restricted, cannot embed here
[...]. Myocardial Perforation by Pacemaker Lead with Insulation Disruption. CTICases. Published August 14, 2015. https://str-case-explorer.pages.dev/c/e3mpz
Imaging findings
Post-procedural chest radiographs showed a change in the position of the right ventricular pacemaker lead compared to intra-procedural images. Corresponding coronal CT confirmed perforation of the right ventricular myocardium by the lead. During attempted lead replacement, the suture sleeve was found to have completely disrupted the insulation and part of the conducting electrode, indicating lead damage.
Key takeaways
Pacemaker lead perforation is a critical complication that can be detected by changes in lead position on serial radiographs or confirmed by CT. Lead damage, such as insulation disruption, can occur during perforation and may necessitate full lead replacement.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- Arterial Central Venous Catheter Placement (into aorta) via IJ and Subclavian VeinIatrogenic
- Retained Suction Catheter Tip (Intraoperative Foreign Body)Iatrogenic
- Choriocarcinoma with Hemorrhagic Pulmonary MetastasesNeoplastic
- Anomalous Left Coronary Artery from Pulmonary Artery (ALCAPA) with Proximal Interruption of the Right Pulmonary ArteryCoronary
- Hypertrophic Cardiomyopathy (HCM)Cardiomyopathy
- Loeffler's Endocarditis (Fibrotic Phase)Cardiomyopathy
- Unilateral Acute Respiratory Distress Syndrome (ARDS) due to Left Pulmonary Artery AgenesisPulmonary Vascular
- Aberrant Right Subclavian ArteryAorta & Arteries
See all cases from August 14, 2015 →
Related Devices & Post-op cases
- Pacemaker Lead Complications (Shredded/Stuck Lead and Broken Lead Fragment)
- Left Atrial Appendage Occlusion Device (Watchman Device)
- CardioMEMS Pulmonary Artery Pressure Monitoring Device
- Azygos Defibrillator Lead in an Azygos Fissure
- Twiddler's Syndrome of an ICD Lead
- Investigational Vagal Nerve Stimulator (CardioFit System) for Heart Failure