Impella Device Failure
[...]. Impella Device Failure. CTICases. Published October 6, 2017. https://str-case-explorer.pages.dev/c/jx2bs
Imaging findings
Serial chest radiographs show an Impella left ventricular assist device that was initially positioned correctly across the aortic valve. A subsequent radiograph shows the catheter has been retracted back into the ascending aorta. Review of the clinical history reveals the device had failed/stopped functioning, and was intentionally pulled back by the cardiologists to avoid impairing native aortic valve function prior to surgical replacement.
Key takeaways
An Impella is a catheter-mounted microaxial flow pump positioned across the aortic valve to support left ventricular output. Retraction of the device into the aorta is typically a sign of accidental dislodgement. However, in the setting of abrupt device failure, intentional retraction into the aorta is performed to prevent the non-functioning pump from obstructing the aortic valve orifice, highlighting the importance of clinical correlation before reporting malposition.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- Hard Metal PneumoconiosisILD
- Vaping-Induced Hard Metal PneumoconiosisILD
- Hypoattenuating Leaflet ThickeningValvular
- IVC Filter Strut Fracture and MigrationIatrogenic
- LVAD Outflow Graft CompressionDevices & Post-op
- Subcutaneous Identification MicrochipArtifact
- Fat Embolism SyndromeIatrogenic
- Pseudallescheria boydii MycetomaInfection
See all cases from October 6, 2017 →