Imaging findings
A bedside chest radiograph demonstrates a sudden, subtle gap/separation in the radiopaque line of an indwelling enteral feeding tube. Fluoroscopy confirms complete separation and rupture of the feeding tube. The retrieved tube demonstrates localized balloon-like expansion and a longitudinal burst/rupture of the tube wall.
Key takeaways
Forceful injection of fluids using a syringe (such as a 10cc syringe) to clear an obstructed or clogged feeding tube can generate extremely high intraluminal pressure. This pressure can cause localized ballooning and catastrophic rupture/separation of the tube wall, requiring fluoroscopic snare retrieval of the distal fragment.
AI-assisted summary — may contain errors. Verify against the source video. Learn more
More from this webinar
- Solitary Fibrous Tumor of the PleuraNeoplastic
- Chronic Vaping-Induced Lung FibrosisExposure
- ROR1 CAR T-Cell Therapy-Induced Cytokine Release SyndromeInflammatory
- Foreign Body Granuloma from BioSentry Biopsy Tract SealantInflammatory
- Circumflex Aortic ArchAorta & Arteries
- Double Aortic Arch with Atretic Left Segment and Secundum ASDAorta & Arteries
- Pulmonary HistoplasmosisInfection
- Ruptured Post-Operative Aortic Root PseudoaneurysmAorta & Arteries
See all cases from April 30, 2021 →
Related Iatrogenic cases
- CellCept-Induced Pneumatosis Intestinalis
- Pulmonary Lymphedema secondary to Sarcoid-like Lymphadenopathy (Humira-Induced)
- Iatrogenic Pulmonary Infarction after Metastasectomy
- Iatrogenic Left Hemidiaphragm Injury with Gastrothorax
- Alpha-1 Antitrypsin Deficiency treated with Endobronchial Valve Lung Volume Reduction
- Acute Lung Allograft Rejection